HT-3500S-RF:

SKU: 00-HT3500S-RF

Hipot Tester
2500Vac @ 20mA
3500Vdc @ 5mA
Adjustable Ground Continuity
Portable and Rugged!

$2,590.00

Portable Hipot Tester encased in a padded shipping enclosure with all controls accessible without removing it from the enclosure.

Relay Isolated Interface: See manuals for details

Options

Hipot Tester <br />2500Vac @ 20mA <br />3500Vdc @ 5mA <br />Adjustable Ground Continuity <br />Portable and Rugged! $2,590.00

Designed & Built in the USA

  • 1-Year Warranty
  • 100% Secure Checkout
Guaranteed Safe Checkout

Compliance West introduces our Portable & Rugged Hipot Tester, the HT-3500S-RF.  It is encased in a shipping enclosure roomy enough for cables or a DVM, with all controls accessible without removing it from the enclosure.  This rugged hipot comes with an adjustable ground continuity test in range of 0.08-1.50Ω, and Auto-offset for test leads up to 1.00Ω. Performs 2500V AC or 3500V DC Hipot Test. Neutral/Hot to Ground Leakage current test up to 20mA AC or 5mA DC. Easy to use and connect. Ships complete with cables, manual, and a Calibration Certificate. Made in the USA. Calibration and repair services available. Not sure – contact us!

Features

  • High precision leakage current, duration and ground continuity settings.
  • Allows no load setup of leakage current and test dwell times.
  • Once entered, ground continuity setting, leakage current and test dwell times can be referenced on the front panel display when unit is not testing.
  • Leakage currents can be set from 1-20mA AC or 1-5mA DC in 0.1mA increments.
  • Dwell times can be set from 1-60 seconds in 1 second increments.
  • Two tests with one setup. Conducts UL/CSA Ground continuity and Hipot tests with one setup, so no additional equipment is needed.
  • Build in a portable case 18.6″ wide x 8.4″ high x 16.3″ deep. 22 lbs. Rugged enough to test manufactured housing.
  • Ground continuity check is defeatable for testing two-wire products.
  • Easy equipment connections. Plug your equipment right into the NEMA plug on the front panel. No adapters needed!
  • Optional J2 and J3 cords allow connection of pigtail power leads.
  • Simple test outcome judgement. Green lights mean pass – red lights mean fail. Failing tests sound a buzzer as well.
  • Ready for your UL Inspector out of the box. NIST Traceable Calibration Certificate to MIL-STD45662-A is included. Cal cert is in accordance with UL guidance.
  • Operation Instructions are printed on the rear panel for easy reference.
  • In-depth testing. Both true Arc Dectection and Leakage Current circuits test the product to find problems.
  • RED lights and a buzzer leave no question if a failure has occurred.
  • Easy shift-by-shift Function Test. Compatible with our Patented function checker, the HTT-1and HTT-1S (With 0.1 ohm Ground check Pass/Fail Setup).

Specifications

Documents