MegaPulse 1.2×50-12PF 39Ω, 400-12kV, 3s charge time, 39Ω virtual impedance. In accordance with the requirements in IEC 1180-1 Section 6.2 and other Standards by reference. It is acceptable for use with unpowered samples requiring a 39Ω impedance, with no current waveform defined. It is designed to test IEC 60335, Table 7 in its entirety. This tester is in accordance with the measurement requirements of CTL DSH 251B. The switch is our IER Relay model. The meter shows the peak of the output waveform into an open circuit. Impedance of the device being tested will attenuate this peak value. Pulse delivery one per minute for pulses above 6kV; two per minute for pulses below 6kV. Please see Manuals Section below to view waveforms. This tester can be provided with Option 950-39, which allows testing to IEC 60950 Annex N/IEC 62368-1 Annex D; Line 2. For automated pulse delivery, please order Option TMM-PF for computer control.
MegaPulse 1.2x50-12PF 39:
SKU: 00-MP1.2x50-12PF 39Surge Tester 1.2×50-12PF 39Ω, 400-12kV, 3s charge time, 42Ω conventional impedance
Call for Price